Perkin Elmer 1 TGA system

Back to the list

Perkin Elmer Pyris 1 TGA-FTIR-GC/MS instrument; a hyphenated system with a FT-IR spectrophotometer, a gas chromatograph (GC), and a mass selective spectrometer (MS)

 Hypphen2.jpg

Parts of the hyphenated system: (A) Perkin Elmer SQ 8 S mass spectrometer, (B) Perkin Elmer Clarus 580 gas chromatograph, (C) Perkin Elmer Frontier FT-IR spectrophotometer, (D) Perkin Elmer TL9000 control unit & transfer line & (E) Perkin Elmer Pyris 1 TGA thermogravimetric analyzer. (Picture taken by Saku Mattila.)

Name and Model (Year of Manufacture / Installation)

Perkin Elmer Pyris 1 TGA (2014/2014) &
Perkin Elmer Frontier (2014/2014) &
Perkin Elmer Clarus 580 (2014/2014) &
Perkin Elmer SQ 8 S (2014/2014)

General Information

A hyphenated system including a thermogravimetric analyzer (TGA) instrument witha Fourier Transform-infrared (FT-IR) analyzer, a gas chromatograph (GC) and a mass selective spectrometer (MS).

The thermogravimetric analyzer measures the weight loss of samples upon heating. The device can typically perform the following measurements: thermal loss, decomposition studies, weight % loss. With FT-IR and MS instruments it’s possible to combine other measurements with the thermogravimetric analysis. (=> evolved gas analysis)

Key Specifications and Features

Parts of the system:

  • Perkin Elmer Pyris 1 TGA thermogravimetric analyzer
  • Perkin Elmer TL9000 TG-IR-GCMS interface (control unit & transfer line)
  • Perkin Elmer Frontier FT-IR spectrophotometer
  • Perkin Elmer Clarus 580 gas chromatograph (GC) unit
  • Perkin Elmer SQ 8 S mass spectrometry (MS) instrument

PE1TGA.jpg

Perkin Elmer Pyris 1 TGA instrument. (Picture on the right Saku Mattila.)


Perkin Elmer Pyris 1 TGA

Sample mass range: from 1 to about 50 mg

Sample Atmosphere: Static or dynamic, including nitrogen or air. Analyses are done at normal pressures.

Sample pans

  • Platinum or Ceramic with capacity of 60 μL

Standard furnace

  • Temperature range: Subambient to 1000 °C
  • Scanning rates: 0.1 to 100 °C/min
  • Temperature precision: ± 2 °C

 

Cooling

  • Forced air cooled with an external fan and internal booster purge
  • Standard Furnace: 1000 °C to 40 °C in less than 15 minutes under normal operation

Balance

  • Tare: Reproducible to ±2 μg
  • Sensitivity: 0.1 μg
  • Accuracy: Better than 0.02 %
  • Precision: 0.001 %

TGA design: A vertical design with a high sensitivity balance and quick response furnace. The balance is located above the furnace and is thermally isolated from it. A precision hang-down wire is suspended from the balance down into the furnace. At the end of the hang-down wire is the sample pan. The sample pan’s position is reproducible.

 

PEftir.jpg 

Perkin Elmer Frontier FT-IR spectrometer and PE TL 9000 control unit. (Picture: Saku Mattila.)

Perkin Elmer Frontier FT-IR spectrometer

  • Wavelength range: 8300 - 350 cm-1 optimized, proprietary KBr beamsplitter
  • Spectral resolution:  0.4 cm-1 for the 3028 cm-1 band in Methane
  • Wavenumber Precision: 0.008 cm-1 at 2000 cm-1
  • Wavenumber Accuracy: 0.02 cm-1 at 2000 cm-1

Optical system:

  • Sealed and desiccated optical unit. Vibration isolated baseplate.
  • DTGS detector
  • Beamsplitter: Proprietary extended range KBr

 PESQ8S.jpg

Perkin Elmer SQ 8S mass selective spectrometer. (Picture: Mika Pirttimäki)

Perkin Elmer SQ 8S mass spectrometer

  • Ion source: Perkin Elmer SMARTsource TM 
  • Filament: Marathon TM
  • Mass range / Resolution: 1.0 - 1200 u (amu) / 0.5 m/z
  • Detector: Clarifi TM detector (a high sensitivity electron multiplier detector)
  • Analyzer: Quadrupole with pre-filter
  • Mass stability: ± 0.1 m/z over 48 hours
  • EI voltage: 10 - 100 eV
  • Temperature settings: SMARTsource: settable 50 °C to 350 °C, GC transfer line: settable 50 °C to 350 °C
  • Scan rate:  up 12500 amu / sec (fully variable)
  • Acquisition rate: 100 points / sec
  • Functions per run: 32 functions or 32 ions per function
  • Linear dynamic range: 106 dependant on acquisition rate
  • TurboMass 6.0 software

 Clarus580.jpg

Perkin Elmer Clarus GC instrument. (Picture: Saku Mattila.)

Perkin Elmer Clarus 580 gas chromatograph

  • Used components of the instrument: the column oven and the silica column
  • Temperature range: 10 °C above ambient to 450 °C
  • Column overheat protect: User settable up to 450 °C
  • Temperature programmer: 3 ramp, 4 plateaus
  • Initial time: 0 to 999 min, increment 0.1 min
  • Rate: 0.1 to 45 °C/min, increment 0.1 °C
  • Plateau time: 0 to 999 min, increment 0.1 min

 

Location, Contact Person

Department of Chemistry, YE523 Manu Lahtinen