Waters 996 HPLC

Back to the list

Waters 996 High Performance Liquid Chromatography (HPLC) system

Waters996.jpg

 Waters 996 HPLC system. (Picture taken by Saku Mattila.)

Name and Model (Year of Manufacture / Installation)

Waters 996 HPLC system (?/?)

General Information

A high performance liquid chromatography (HPLC) system. The equipment is used for analyzing and characterizing the compounds. The system can perform standard liquid chromatography analysis and also size-exclusion chromatography (SEC) (or gel permeation chromatography (GPC)).

Key Specifications        

Parts of the system:

  • Waters 996 Liquid Chromatography unit
  • Waters 996 Photodiode Array (PDA) Detector
  • Waters 410 Differential Refractometer
  • Waters System Interface Module
  • Polymer Laboratories PL-ELS-1000 Evaporative Light Scattering Detector
  • Waters 717plus Autosampler
  • Waters Temperature Control Module
  • 2 pumps

Waters 996 Phodiode Array Detector

 

Wavelength range: 190 to 800 nm
Wavelength accuracy: ± 1 nm
Linearity range (Per ASTM 685-79): 5 % at 2.0 AU (probylparaben at 258 nm)
Spectral resolution: 1.2 nm
Baseline noise: ± 1.5 x 10-5 AU peak-to-peak, dry (at 254 nm)
Drift: 1 x 10-3 AU/hour (at 254 nm, after warmup)

Flow cells                        Path length (mm)               Tube inner diameter (inch)

Standard                             10                                       0.009 (0.23 mm)                
Semi-preparative                 3                                         0.040 (1.016 mm)
Microbore                            3                                         0.005 (0.127 mm)
Inert                                    10                                        0.001 (0.254 mm)

Polymer Laboratories PL-ELS-1000 Evaporative Light Scattering Detector

 

  • Light source: Tungsten/Halogen lamp
  • Detector: Photodiode

Temperature range

  • Evaporator: 30 - 300 °C
  • Nebulizer: 30 - 220 °C

Nebulizer gas

  • Typical inlet pressure: 413 - 689 kPa (60 - 100 psi)
  • Maximum inlet pressure: 1 MPa (150 psi)
  • Gas flow rate: 0 - 1.0 SLM (standard litre per minute)
  • Eluent flow rate: 0 - 0.5 mL/min  
  • Analog output:  0 - 1 V and 0 - 10 V FSD

Waters 410 Differential Refractometer

 

Rl Range: 1.00 to 1.75 RIU

Measurement Range:
5x10-3 RIU maximum FS (SENS= 1, SF=1)
5 x 10-a RIU minimum FS (SENS=1024, SF=100)

Flow Rate: 0.3 - 10 mL/min

Noise: 2.0 x 10-8 RIU, (TC = 1, SF = 20, SENS = 32, THF at 24 °C with restrictor coil)

Drift: 2.5 x 10-7 RlU/hour (static with THF)

Sensitivity Settings: 1, 2, 4, 8, 16, 32, 64, 128, 256, 512, 1024

Time Constant Filter Settings: 0.2, 1, 3, 10 seconds

Recorder Output: 0 to 10 mV

Temperature Control:  

  • Internal oven: 30 to 50 °C
  • External column heaters: Ambient to 150 °C

Integrator output

Sensitivity setting                      Output (V)      

1                                                      1
2                                                      2
4                                                      4
8                                                      8
16 to 1024                                      10


Optical Component Specifications


Light Source: 930 nm pulsed Light Emitting Diode (LED)

Flow Cell: Fused quartz

Cell Volume: 10 μL

Pressure Limit: 689 kPa (100 psi)

Fluid Path Materials: 316 stainless steel, Dynasii®, Suprasil II®, Teflon®, Kalrez®, quartz.

Key Features

PL-ELS-1000 Evaporative Light Scattering Detector:

+ PL-ELS 1000 control software enables:

  • temperature and gas flow control
  • signal monitoring
  • methods storage
Location, Contact Person

Department of Chemistry, YSK338 / Hannu Pakkanen